PAROX 1200
Measuring principle
MBE – dedicated to successful partnerships
Technical specifications
PAROX 1200 U unheated cell with no compensation
PAROX 1200 H heated cell with temperature compensation
Measuring range min.
Measuring range max.
Measuring range max.
0 – 5% O2
0 – 100% O2
0 – 100% O2
0 – 5% O2
0 – 100% O2 / 95 – 100% O2
0 – 100% O2 / 95 – 100% O2
Power supply
5V DC – 28V DC
max. 1,5W
max. 1,5W
12 – 28V DC
max. 12W
max. 12W
Output value
- Output value (Option)
- galvanic isolated
- Output value (Option)
- galvanic isolated
0 – 1V or 0 - 4V
4 – 20 mA or 0 – 10V
yes
4 – 20 mA or 0 – 10V
yes
0 – 1V or 0 – 4V
4 – 20 mA or 0 – 10V
yes
4 – 20 mA or 0 – 10V
yes
Response time
T90 at 150 ml/min
gas flow
Max. gas flow
T90 at 150 ml/min
gas flow
Max. gas flow
< 3s
300 ml/min.
300 ml/min.
< 3s
300 ml/min.
300 ml/min.
Operating temperature
15 – 30 °C
0 – 60 °C with customer’s own compensation
0 – 60 °C with customer’s own compensation
0 – 45 °C
Heating temperature
55 °C
Temperature sensor
optional
inbuilt
Typical performance under constant ambient conditions
Linearity
< ± 0,1% O2
< ± 0,1% O2
Zero setting deviation
Repeatability
Repeatability
< 0,1% O2 / week
< ± 0,05% O2
< ± 0,05% O2
< 0,1% O2 / week
< ± 0,02% O2
< ± 0,02% O2
Temperature deviation
- zero drift
- sensitivity drift
- zero drift
- sensitivity drift
< ± 0,1 % O2 / °C
< ± 1 % of measured
value / °C
< ± 0,03 % O2 / °C
< ± 0,1 % of measured
value / °C
Mechanical dimensions PAROX 1200
Price list and operating manual by E-Mail on request.